| Reliable stability: |
¡ñAdopting the special designed air pump,it provides users stable 2-20hpa therapy pressure
¡ñRamp capabilities allow a 0-60 minutes period with adjustable ramp starting pressure
¡ñInspiration trigger for auto start-up
¡ñAutomatic leakage compensation and 4-level altitude compensation ensure the therapy accuracy anywhere
¡ñIn time alarm function when mask and/or tubing off-line alarm when accidental power-off Ergonomic design
¡ñ Lighted user buttons with lockup capability
¡ñIntegrated designed InH2 TM heated humidifier
¡ñEmbedded smart memory records usage Blog
¡ñQuiet operation |
Technical Specification: |
Device size
|
Dimensions: |
220¡Á94¡Á12mm |
| 313¡Á94¡Á12mm(with InH2 TM heated humidifier) |
Weight:©‚2Kg |
Standard compliance |
IEC 60601-1 General Requirements for safety of Medical Electrical Equipment;
ISO 17510 Sleep Aponoea Breathing Therapy |
Mode of Operation
|
Continuous |
AC power Consumption
|
100-240VAC,50/60Hz |
Type of Protection Against Electric Shock
|
Class II Equipment |
Degree of Protection Againest Electric Shock
|
Type BF Applied Part |
Degree of Protection Againest Ingtess of Water |
IPXO-Ordinary Equipment |
Pressure Range |
2 to 20 hpa(0.5 hpa increments) |
Pressure Stability |
2 to 20 hpa(¡À0.5 hpa) |
Maximun Flow
|
>15LPM |
Pressure Display Accy
|
¡À0.5 hpa |
Sound level
|
¡Ü27dB |